For making the observations you need (besides this instruction):
    
    
      Every few hours and always with a calibration source (before or after)
      you must do the calibration of the noise diode ("CAL"). Check the error
      and if it is large (several mV), there is something wrong with the
      conditions, so check the weather or make a new CAL-observation.
    
The primary calibration source DR21 must be observed many times during
      an observation shift, and always when the conditions change (if it can be
      observed, if not, use secondary calibrators instead). You can observe it
      between 14.30-18.30 and 22.30-01.30 LST, both times 1-2 times. At other
      times you choose the suitable one from 3C84 and 3C274.
      
      You must get a good result from a calibration observation, so if the
      pointing is not good, the fluxlevel is low or error is large, all
      observations with that calibration must be thrown away. So, if you don't
      get good results from a calibration observation, do not continue
      observations until you do get a good result! (unless there is no
      calibration source available and you could observe one after an hour or
      so, you can continue observations and hope that conditions stay the
      same). If you have stopped the observations because of bad weather, and
      when the conditions get better, are thinking of starting them again, it
      is good to start by checking the fluxlevels of a calibration source.
      
      "A good fluxlevel" depends on the sources elevation and other conditions.
      You can check earlier results from the log book and then see what is a
      normal fluxlevel. If the weather is good and the fluxlevel of a
      calibration source is low, check the offsets and humidity / heater (see
      next paragraph).
      
    
The heater is used in winter to melt snow from the radome and dry radome
      after rain. In summer to dry radome after rain and remove dew from the
      radome.
      
      In summer: dew is falling on the radome during night time, and that
      lowers the fluxlevels. Even if the weather seems perfect, there will
      almost always be some dew, and you must try to prevent that. So in the
      evening when the humidity starts to rise to about 80% you start to use
      the heater (position 1 from the actual heater and then position "talvi"
      from the main switch). Check instructions for using heater and autoheat
      from the instruction list.
      You keep it on about 15-20 min and then off about 40-45 min (depending on
      the humidity). Check the fluxlevels to see if they are changing. If they
      are rising a lot it might be wise to keep the heater on a bit longer
      until the flux levels have stabilised. And when the heater is used it is
      extremely important to write down the temperatures (expecially Tant)
      because they can change rapidly. After rain the heater is kept on until
      the radome is (hopefully) dry and then after the temperature has
      normalized start the observations. It is wise to observe calibration
      source when you are drying the radome, so you know when the flux levels
      are normal again.
      
      In Winter: Drying the radome after rain similarily as above, but also
      melting the snow. Usually you melt the snow with "full power", so the
      settings are position 2 from the heater and position "Talvi" / "Winter"
      from the main switch. When the weather is cold it can take time to melt
      the snow, and you must check the radome if the snow has really melted and
      is there water dripping from it. After melting the snow it can take hours
      to the temperature to be normalized again and you can start observing.
      And when it's really cold (-10C or more) it is almost impossible to melt
      the snow.
      
    
Last modified: Fri Dec 19 16:14:12 EET 2008